Difference between revisions of "Team:Hong Kong-CUHK/InterLab"

Line 30: Line 30:
 
     border: 1px solid black;
 
     border: 1px solid black;
 
}
 
}

 </style>  
+
 
 +
p .main {
 +
    text-align: left;
 +
}
 </style>  
 
   </head>
 
   </head>
  
Line 40: Line 43:
 
       </div>
 
       </div>
 
     </div>
 
     </div>
<div class="column full_size">
 
  
  
  
<div class="column full_size judges-will-not-evaluate">
 
<h3>★  ALERT! </h3>
 
<p>This page is used by the judges to evaluate your team for the <a href="https://2017.igem.org/Judging/Medals">medal criterion</a> or <a href="https://2017.igem.org/Judging/Awards"> award listed above</a>. </p>
 
<p> Delete this box in order to be evaluated for this medal criterion and/or award. See more information at <a href="https://2017.igem.org/Judging/Pages_for_Awards"> Instructions for Pages for awards</a>.</p>
 
</div>
 
 
<div class="clear"></div>
 
<div class="clear"></div>
  
 
<div class="column full_size">
 
<div class="column full_size">
<h1>InterLab</h1>
 
<h3>Bronze Medal Criterion #4</h3>
 
<p><b>Standard Tracks:</b> Participate in the Interlab Measurement Study and/or improve the characterization of an existing BioBrick Part or Device and enter this information on that part's Main Page in the Registry. The part that you are characterizing must NOT be from a 2017 part number range.
 
<br><br>
 
For teams participating in the <a href="https://2017.igem.org/Competition/InterLab_Study">InterLab study</a>, all work must be shown on this page.
 
  
 
<h3>Background</h3>
 
<h3>Background</h3>
Line 205: Line 197:
 
<br>
 
<br>
 
</p>
 
</p>
 +
<br>
 +
<br>
  
 +
<h3>Experiment</h3>
 +
<p class="main">
 +
The colonies was observed from blue light box.
 +
The controls show that the plasmid containing GFP gene will give fluorescence.
 +
<br>
 +
<br>
 +
The strength of promoter is correlated to the strength of promoter. The higher the strength, the higher the expression of GFP. Among the colonies from devices 4-6, the colonies from device 4 show the highest fluorescence while those from device 6 show the lowest.
 +
Among the colonies from devices 1-3, the colonies from device 3 show the lowest fluorescence but <font color="red">no significant difference</font> in fluorescence between the colonies from devices 1 and 2.
 +
</p>
  
</div>
+
<p><img src="https://static.igem.org/mediawiki/2017/1/12/Cuhkigemcolony.PNG"  style="width:400px;height:500px;" ></p>
  
 +
<p><img src="https://static.igem.org/mediawiki/2017/9/9c/CuhkigemOD600graph.PNG"  style="width:800px;height:400px;" ></p>
 +
<br><br>
  
 +
<p>
 +
<b>Fig.1 </b>shows an increasing trend of the cell growth under OD600 absorbance and Fig.2 shows an increasing trend of the fluorescence. <br>
 +
<b>Fig.3 </b> shows the trend of fluorescence per cell.<br>
 +
Among devices 4-6, the data match with the fluorescence of their colonies and the strength of their promoters. Device 4 has the highest fluorescence per cell and device 6 has the lowest. Among devices 1-3, the data also match with the fluorescence of their colonies and the strength of their promoters. Device 1 has the highest fluorescence per cell and device 3 has the lowest.
 +
<br><br>
 +
Interestingly, the fluorescence per cell of device 6 and negative control are very close. The fluorescence per cell of device 6 is even lower than the negative control at some time points.
 +
This shows that device 6 produces very little amount of GFP and the result may also be affected by the measurement error in plate reader.
 +
<br><br>
 +
Another interesting point is the fluorescence per cell of all devices peaked at (t=2hours). One of the possible reasons is the cells spend more energy on cell division but not on producing GFP.
 +
</p>
  
 +
<p><img src="https://static.igem.org/mediawiki/2017/e/e9/Cuhkigemfluorescence.PNG"  style="width:500px;height:300px;" ></p>
  
 +
<br><br><br><br>
  
 +
<p><img src="https://static.igem.org/mediawiki/2017/0/03/Cuhkigemfluoresceinconc.PNG"  style="width:500px;height:300px;" ></p>
 +
 +
 +
<br>
 +
<br>
 +
<br>
 +
 +
<p>
 +
Reference<br>
 +
1. Mutalik VK, Guimaraes JC, Cambray G, Lam C, Christoffersen MJ, Mai QA, Tran AB, Paull M, Keasling JD, Arkin AP, Endy D. Precise and reliable gene expression via standard transcription and translation initiation elements. Nat Methods. 2013 Apr;10(4):354-60.
 +
 +
</p>
 +
 +
</div>
  
  
 
</html>
 
</html>
 +
{{Template:Hong_Kong-CUHK_footer}}

Revision as of 12:35, 22 September 2017





Background

Reliable and repeatable measurement is the golden rule of engineering, and so do synthetic biology. However, most of the fluorescent measurement data generated nowadays can not be compared, because fluorescence data are usually reported in relative unit, but not in absolute unit. In addition, different groups may perform measurement with different protocol, which makes it hard to reproduce. Therefore, iGEM develop a green fluorescent protein (GFP) measurement protocol in order to produce a more reliable, repeatable measurement of GFP. GFP is one of the most commonly used reporter for measurement and easily to be measured in most of laboratories. In the protocol,the unit for fluorescence data is unified so that the results can be compared. The InterLab protocol also unifies the measurement procedure and prevents different data processing for the measurement.

The Fourth InterLab

This year, iGEM invited all teams among the world to join the fourth InterLab Study. The aim of the study is to find out how close can the numbers be when fluorescence is measured all around the world using the same InterLab protocol. We registered for the interlab study and measured all the interlab parts using the InterLab plate reader protocol.

Experiment

iGEM provided 8 plasmids for the InterLab Study. Devices 1-6 and positive control have the same reporter gene (GFP), terminator (B0015) and backbone (pSB1C3). However, devices 1-3 share the same RBS (B0034), while devices 4-6 share another modified RBS called bicistronic device (BCD2). Different promoters are also used in different plasmid. According to the strength of promoter described by iGEM2006_Berkeley team, device 1 should have the strongest fluorescence and device 3 should have the weakest among devices 1-3, while Device 4 should have the strongest fluorescence and device 6 should have the weakest among devices 4-6.
Summary of parts measured
Promoter (strength) RBS Coding Sequence Terminator Backbone
Device 1 (J364000) J23101(1791) B0034 GFP B0015 pSB1C3
Device 2 (J364001) J23106(1185)
Device 3 (J364002) J23117(162)
Device 4 (J364003) J23101(1791) BCD2
Device 5 (J364004) J23106(1185)
Device 6 (J364005) J23117(162)
Positive control (I20270) J23151(N/A) B0032
Negative control (R0040) R0040 (N/A) N/A

What is Bicistronic Device (BCD)?

Bicistronic device (BCD) is a modified ribosome binding site (RBS) with another cistron. The device consists of another cistron (cistron 1) with another RBS (SD2) between RBS (SD1) and gene of interest (cistron 2). Also, the stop codon of cistron 1 overlaps the start codon of cistron 2. Ribosome binding efficiency and translation rate will be affected after the secondary structure near the RBS has changed due to the change of gene of interest. This device can maintain the ribosome binding efficiency and translation rate even though the gene of interest has changed. Therefore, it is used to control the amount of fluorescence in this study. BCD is expected to generate a more reliable and precise gene expression.

Method

We follow exactly the interlab protocol provided by iGEM(link) with the below plate reader setting:
Microplate used: Corning® clear flat bottom black 96 well plates
Instrument used: BMG LABTECH’s CLARIOstar®

Instrument Settings for measuring OD600 of LUDOX and Cells:
- Endpoint settings
No. of flashes per well: 10
Scan mode: orbital averaging
Scan diameter [mm]: 3

- Optic settings Excitation: 600

- General settings
Top optic used
Settling time [s]: 0.1
Reading direction: bidirectional, horizontal left to right, top to bottom
Target temperature [°C]: 25

Instrument Settings for measuring fluorecence of Fluorescein and GFP:
- Endpoint settings
No. of flashes per well: 8
Scan mode: orbital averaging
Scan diameter [mm]: 3

- Optic settings
Excitation: 470-15
Emission: 515-20
Gain: 500
Focal height [mm]: 9

- General settings
Top optic used
Reading direction: bidirectional, horizontal left to right, top to bottom
Target temperature [°C]: 25



Experiment

The colonies was observed from blue light box. The controls show that the plasmid containing GFP gene will give fluorescence.

The strength of promoter is correlated to the strength of promoter. The higher the strength, the higher the expression of GFP. Among the colonies from devices 4-6, the colonies from device 4 show the highest fluorescence while those from device 6 show the lowest. Among the colonies from devices 1-3, the colonies from device 3 show the lowest fluorescence but no significant difference in fluorescence between the colonies from devices 1 and 2.



Fig.1 shows an increasing trend of the cell growth under OD600 absorbance and Fig.2 shows an increasing trend of the fluorescence.
Fig.3 shows the trend of fluorescence per cell.
Among devices 4-6, the data match with the fluorescence of their colonies and the strength of their promoters. Device 4 has the highest fluorescence per cell and device 6 has the lowest. Among devices 1-3, the data also match with the fluorescence of their colonies and the strength of their promoters. Device 1 has the highest fluorescence per cell and device 3 has the lowest.

Interestingly, the fluorescence per cell of device 6 and negative control are very close. The fluorescence per cell of device 6 is even lower than the negative control at some time points. This shows that device 6 produces very little amount of GFP and the result may also be affected by the measurement error in plate reader.

Another interesting point is the fluorescence per cell of all devices peaked at (t=2hours). One of the possible reasons is the cells spend more energy on cell division but not on producing GFP.








Reference
1. Mutalik VK, Guimaraes JC, Cambray G, Lam C, Christoffersen MJ, Mai QA, Tran AB, Paull M, Keasling JD, Arkin AP, Endy D. Precise and reliable gene expression via standard transcription and translation initiation elements. Nat Methods. 2013 Apr;10(4):354-60.