+ | <h1>INTERLAB STUDY</h1> | ||
+ | <div class="main"> | ||
+ | <p> | ||
+ | We participated in the Fourth International InterLab Measurement Study. They asked us to measure the fluorescence of GFP in different devices in an attempt to explore the question, “How close can the numbers be when fluorescence is measured all around the world? “. | ||
+ | <br /> We followed the same protocols as directed in the iGEM protocol page. | ||
− | + | </p> | |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | < | + | <h3> |
− | + | Materials | |
− | + | </h3> | |
− | <p> | + | <p> |
− | + | E.coli DH5-alpha: Competent cells were purchased from TOYOBO<br /> | |
− | + | Plate reader: SH-9000Lab | |
− | </p | + | </p> |
− | + | ||
+ | <h3>Result</h3> | ||
+ | |||
+ | <h4>1. Calculation</h4> | ||
+ | <table> | ||
+ | <caption> | ||
+ | Table 1. Abs600 of LUDOX-S40 and H<sub><small>2</small></sub>O | ||
+ | </caption> | ||
+ | <thead> | ||
+ | <td></td> | ||
+ | <td>LUDOX-HS40</td> | ||
+ | <td>H<sub><small>2</small></sub>O</td> | ||
+ | </thead> | ||
+ | <tbody> | ||
+ | <tr> | ||
+ | <th>Replicate1</th> | ||
+ | <td>0.034</td> | ||
+ | <td>0.025</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate2</th> | ||
+ | <td>0.031</td> | ||
+ | <td>0.024</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate3</th> | ||
+ | <td>0.032</td> | ||
+ | <td>0.025</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate4</th> | ||
+ | <td>0.031</td> | ||
+ | <td>0.023</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Arith.Mean</th> | ||
+ | <td>0.032</td> | ||
+ | <td>0.023</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Corrected Abs600</th> | ||
+ | <td>0.00775</td> | ||
+ | <td></td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Reference OD600</th> | ||
+ | <td>0.0425</td> | ||
+ | <td></td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>OD600/Abs600</th> | ||
+ | <td>5.483871</td> | ||
+ | <td></td> | ||
+ | </tr> | ||
+ | </tbody> | ||
+ | </table> | ||
+ | |||
+ | |||
+ | <h4>2. FIuorescein fluorescence standard curve</h4> | ||
+ | <p>Table.2 shows the fluorescence measurement data Fig.1 shows the standard curve. We couldn’t measure the value of fluorescein fluorescence at 50, 25, 12.5, 6.25, 3.125 μM because they exceeded the measurable range.</p> | ||
+ | |||
+ | <table> | ||
+ | <caption> | ||
+ | Table 2. fluorescence data for fluorescein standard curve | ||
+ | </caption> | ||
+ | <tr> | ||
+ | <th>uM Fluorescein</th> | ||
+ | <th>1.5625</th> | ||
+ | <th>0.78125</th> | ||
+ | <th>0.390625</th> | ||
+ | <th>0.195313</th> | ||
+ | <th>0.097656</th> | ||
+ | <th>0.048828</th> | ||
+ | <th>0</th> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate 1</th> | ||
+ | <td>57809</td> | ||
+ | <td>29980</td> | ||
+ | <td>15540</td> | ||
+ | <td>7861</td> | ||
+ | <td>3913</td> | ||
+ | <td>1949</td> | ||
+ | <td>3</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate 2</th> | ||
+ | <td>58424</td> | ||
+ | <td>30336</td> | ||
+ | <td>15369</td> | ||
+ | <td>7579</td> | ||
+ | <td>3948</td> | ||
+ | <td>1993</td> | ||
+ | <td>3</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate 3</th> | ||
+ | <td>58075</td> | ||
+ | <td>29993</td> | ||
+ | <td>15535</td> | ||
+ | <td>7723</td> | ||
+ | <td>3908</td> | ||
+ | <td>1999</td> | ||
+ | <td>2</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Replicate 4</th> | ||
+ | <td>57953</td> | ||
+ | <td>29694</td> | ||
+ | <td>15532</td> | ||
+ | <td>7798</td> | ||
+ | <td>3935</td> | ||
+ | <td>2044</td> | ||
+ | <td>4</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Arith. Mean</th> | ||
+ | <td>58065.25</td> | ||
+ | <td>30000.75</td> | ||
+ | <td>15494</td> | ||
+ | <td>7740.25</td> | ||
+ | <td>3926</td> | ||
+ | <td>1996.25</td> | ||
+ | <td>3</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>S.D.</th> | ||
+ | <td>262.7171</td> | ||
+ | <td>262.665</td> | ||
+ | <td>83.39864</td> | ||
+ | <td>121.4012</td> | ||
+ | <td>18.77942</td> | ||
+ | <td>38.86194</td> | ||
+ | <td>0.816497</td> | ||
+ | </tr> | ||
+ | |||
+ | </table> | ||
+ | <small>Fig1. Fluorescein standard curve</small> | ||
+ | <img src="https://2017.igem.org/File:T--Kobe--interlab_fig1.jpg" width="80%" /> | ||
+ | |||
+ | <h3>3. Cell measurement</h3> | ||
+ | |||
+ | <table> | ||
+ | <caption>Table 3.Abs600 (cell concentration)</caption> | ||
+ | <tr> | ||
+ | <th></th> | ||
+ | <th>0h</th> | ||
+ | <th>2h</th> | ||
+ | <th>4h</th> | ||
+ | <th>6h</th> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 1)</th> | ||
+ | <td>0.00475</td> | ||
+ | <td>0.0275</td> | ||
+ | <td>0.14125</td> | ||
+ | <td>0.25875</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 2)</th> | ||
+ | <td>0.00975</td> | ||
+ | <td>0.0315</td> | ||
+ | <td>0.13275</td> | ||
+ | <td>0.235</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 1)</th> | ||
+ | <td>0.007</td> | ||
+ | <td>0.01475</td> | ||
+ | <td>0.12125</td> | ||
+ | <td>0.25875</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 2)</th> | ||
+ | <td>0.00825</td> | ||
+ | <td>0.019</td> | ||
+ | <td>0.1235</td> | ||
+ | <td>0.26925</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 1)</th> | ||
+ | <td>0.0045</td> | ||
+ | <td>0.00375</td> | ||
+ | <td>0.00875</td> | ||
+ | <td>0.01</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 2)</th> | ||
+ | <td>0.0045</td> | ||
+ | <td>0.00325</td> | ||
+ | <td>0.009</td> | ||
+ | <td>0.007</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 1)</th> | ||
+ | <td>0.006</td> | ||
+ | <td>0.01825</td> | ||
+ | <td>0.12825</td> | ||
+ | <td>0.2665</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 2)</th> | ||
+ | <td>0.00925</td> | ||
+ | <td>0.02325</td> | ||
+ | <td>0.141</td> | ||
+ | <td>0.26375</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 1)</th> | ||
+ | <td>0.0065</td> | ||
+ | <td>0.02225</td> | ||
+ | <td>0.13425</td> | ||
+ | <td>0.2735</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 2)</th> | ||
+ | <td>0.0105</td> | ||
+ | <td>0.0275</td> | ||
+ | <td>0.144</td> | ||
+ | <td>0.26325</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 1)</th> | ||
+ | <td>0.00375</td> | ||
+ | <td>0.00325</td> | ||
+ | <td>0.00875</td> | ||
+ | <td>0.0445</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 2)</th> | ||
+ | <td>0.0065</td> | ||
+ | <td>0.004</td> | ||
+ | <td>0.0085</td> | ||
+ | <td>0.0125</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 1)</th> | ||
+ | <td>0.00575</td> | ||
+ | <td>0.02375</td> | ||
+ | <td>0.14775</td> | ||
+ | <td>0.28175</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 2)</th> | ||
+ | <td>0.00875</td> | ||
+ | <td>0.019</td> | ||
+ | <td>0.1275</td> | ||
+ | <td>0.27675</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 1)</th> | ||
+ | <td>0.006</td> | ||
+ | <td>0.02425</td> | ||
+ | <td>0.1505</td> | ||
+ | <td>0.2855</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 2)</th> | ||
+ | <td>0.0085</td> | ||
+ | <td>0.02125</td> | ||
+ | <td>0.1005</td> | ||
+ | <td>0.2625</td> | ||
+ | </tr> | ||
+ | </table> | ||
+ | <img src="https://2017.igem.org/File:T--Kobe---interlab_fig2.jpg" /> | ||
+ | <small>Fig.2: Abs600 (cell concentration)</small> | ||
+ | |||
+ | <table> | ||
+ | <caption> Table 4: fluorescence</caption> | ||
+ | <tr> | ||
+ | <th></th> | ||
+ | <th>0h</th> | ||
+ | <th>2h</th> | ||
+ | <th>4h</th> | ||
+ | <th>6h</th> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 1)</th> | ||
+ | <td>39.5</td> | ||
+ | <td>26.25</td> | ||
+ | <td>123</td> | ||
+ | <td>79</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 2)</th> | ||
+ | <td>34.5</td> | ||
+ | <td>15</td> | ||
+ | <td>120</td> | ||
+ | <td>112</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 1)</th> | ||
+ | <td>72.5</td> | ||
+ | <td>316.25</td> | ||
+ | <td>1272.75</td> | ||
+ | <td>2534.25</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 2)</th> | ||
+ | <td>68.25</td> | ||
+ | <td>318.25</td> | ||
+ | <td>1269.25</td> | ||
+ | <td>2440.25</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 1)</th> | ||
+ | <td>557.5</td> | ||
+ | <td>563</td> | ||
+ | <td>940.5</td> | ||
+ | <td>1046.5</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 2)</th> | ||
+ | <td>461.75</td> | ||
+ | <td>355.25</td> | ||
+ | <td>633.75</td> | ||
+ | <td>781.25</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 1)</th> | ||
+ | <td>83.75</td> | ||
+ | <td>374.75</td> | ||
+ | <td>2280.25</td> | ||
+ | <td>5128.5</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 2)</th> | ||
+ | <td>116.25</td> | ||
+ | <td>494.25</td> | ||
+ | <td>2128.75</td> | ||
+ | <td>5528.5</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 1)</th> | ||
+ | <td>3</td> | ||
+ | <td>29.5</td> | ||
+ | <td>154.75</td> | ||
+ | <td>232.75</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 2)</th> | ||
+ | <td>28.5</td> | ||
+ | <td>29.75</td> | ||
+ | <td>161</td> | ||
+ | <td>258.5</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 1)</th> | ||
+ | <td>94</td> | ||
+ | <td>172.5</td> | ||
+ | <td>327.75</td> | ||
+ | <td>769</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 2)</th> | ||
+ | <td>71.75</td> | ||
+ | <td>165.5</td> | ||
+ | <td>286</td> | ||
+ | <td>443.5</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 1)</th> | ||
+ | <td>7.5</td> | ||
+ | <td>196</td> | ||
+ | <td>614.5</td> | ||
+ | <td>858.25</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 2)</th> | ||
+ | <td>14.5</td> | ||
+ | <td>189.75</td> | ||
+ | <td>796.75</td> | ||
+ | <td>987</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 1)</th> | ||
+ | <td>60.25</td> | ||
+ | <td>29.5</td> | ||
+ | <td>129.5</td> | ||
+ | <td>164</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 2)</th> | ||
+ | <td>33.75</td> | ||
+ | <td>26.5</td> | ||
+ | <td>97.75</td> | ||
+ | <td>165</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th></th> | ||
+ | </tr> | ||
+ | </table> | ||
+ | <!--fig3を挿入--> | ||
+ | <img src="https://2017.igem.org/File:T--Kobe--interlab_fig3.jpg" /> | ||
+ | <small>Fig.3: Fluorescence</small> | ||
+ | <table> | ||
+ | <caption>Table 6. Value of Flu/Abs<sub><small>600</small></sub></caption> | ||
+ | <tr> | ||
+ | <th></th> | ||
+ | <th>0h</th> | ||
+ | <th>2h</th> | ||
+ | <th>4h</th> | ||
+ | <th>6h</th> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 1)</th> | ||
+ | <td>0.045689</td> | ||
+ | <td>0.00469</td> | ||
+ | <td>0.004282</td> | ||
+ | <td>0.001499</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Negative Control (Colony 2)</th> | ||
+ | <td>0.018247</td> | ||
+ | <td>0.002376</td> | ||
+ | <td>0.004481</td> | ||
+ | <td>0.002352</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 1)</th> | ||
+ | <td>0.055533</td> | ||
+ | <td>0.105486</td> | ||
+ | <td>0.051542</td> | ||
+ | <td>0.048113</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Positive Control (Colony 2)</th> | ||
+ | <td>0.04131</td> | ||
+ | <td>0.082437</td> | ||
+ | <td>0.050482</td> | ||
+ | <td>0.044579</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 1)</th> | ||
+ | <td>0.607923</td> | ||
+ | <td>0.745125</td> | ||
+ | <td>0.532061</td> | ||
+ | <td>0.520307</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 1 (Colony 2)</th> | ||
+ | <td>0.522086</td> | ||
+ | <td>0.640948</td> | ||
+ | <td>0.355477</td> | ||
+ | <td>0.553925</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 1)</th> | ||
+ | <td>0.070761</td> | ||
+ | <td>0.101285</td> | ||
+ | <td>0.087415</td> | ||
+ | <td>0.094541</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 2 (Colony 2)</th> | ||
+ | <td>0.063885</td> | ||
+ | <td>0.104249</td> | ||
+ | <td>0.074675</td> | ||
+ | <td>0.103012</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 1)</th> | ||
+ | <td>0.001716</td> | ||
+ | <td>0.006526</td> | ||
+ | <td>0.005688</td> | ||
+ | <td>0.004194</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 3 (Colony 2)</th> | ||
+ | <td>0.01537</td> | ||
+ | <td>0.005312</td> | ||
+ | <td>0.005489</td> | ||
+ | <td>0.004827</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 1)</th> | ||
+ | <td>0.118182</td> | ||
+ | <td>0.265189</td> | ||
+ | <td>0.184017</td> | ||
+ | <td>0.084842</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 4 (Colony 2)</th> | ||
+ | <td>0.054166</td> | ||
+ | <td>0.218167</td> | ||
+ | <td>0.166671</td> | ||
+ | <td>0.177319</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 1)</th> | ||
+ | <td>0.00281</td> | ||
+ | <td>0.040555</td> | ||
+ | <td>0.020409</td> | ||
+ | <td>0.014984</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 5 (Colony 2)</th> | ||
+ | <td>0.007484</td> | ||
+ | <td>0.049747</td> | ||
+ | <td>0.030945</td> | ||
+ | <td>0.017517</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 1)</th> | ||
+ | <td>0.050096</td> | ||
+ | <td>0.005969</td> | ||
+ | <td>0.004225</td> | ||
+ | <td>0.002821</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th>Device 6 (Colony 2)</th> | ||
+ | <td>0.024278</td> | ||
+ | <td>0.006149</td> | ||
+ | <td>0.004835</td> | ||
+ | <td>0.003078</td> | ||
+ | </tr> | ||
+ | <tr> | ||
+ | <th></th> | ||
+ | </tr> | ||
+ | </table> | ||
+ | <small>Fig.4; Value of Flu/Abs<sub><small>600</small></sub></small> | ||
+ | <img src="https://2017.igem.org/File:T--Kobe--interlab_fig4.jpg" /> | ||
+ | |||
+ | <style> | ||
+ | table { | ||
+ | border-collapse: collapse; | ||
+ | text-align: left; | ||
+ | line-height: 1.5; | ||
+ | } | ||
+ | |||
+ | table th { | ||
+ | width: 100px; | ||
+ | padding: 10px; | ||
+ | font-weight: bold; | ||
+ | vertical-align: top; | ||
+ | border: 1px solid #ccc; | ||
+ | } | ||
+ | |||
+ | table td { | ||
+ | padding: 10px; | ||
+ | vertical-align: top; | ||
+ | border: 1px solid #ccc; | ||
+ | } | ||
+ | </style> | ||
+ | |||
+ | </div> | ||
Revision as of 05:46, 27 September 2017
INTERLAB STUDY
We participated in the Fourth International InterLab Measurement Study. They asked us to measure the fluorescence of GFP in different devices in an attempt to explore the question, “How close can the numbers be when fluorescence is measured all around the world? “.
We followed the same protocols as directed in the iGEM protocol page.
Materials
E.coli DH5-alpha: Competent cells were purchased from TOYOBO
Plate reader: SH-9000Lab
Result
1. Calculation
LUDOX-HS40 | H2O | |
Replicate1 | 0.034 | 0.025 |
---|---|---|
Replicate2 | 0.031 | 0.024 |
Replicate3 | 0.032 | 0.025 |
Replicate4 | 0.031 | 0.023 |
Arith.Mean | 0.032 | 0.023 |
Corrected Abs600 | 0.00775 | |
Reference OD600 | 0.0425 | |
OD600/Abs600 | 5.483871 |
2. FIuorescein fluorescence standard curve
Table.2 shows the fluorescence measurement data Fig.1 shows the standard curve. We couldn’t measure the value of fluorescein fluorescence at 50, 25, 12.5, 6.25, 3.125 μM because they exceeded the measurable range.
uM Fluorescein | 1.5625 | 0.78125 | 0.390625 | 0.195313 | 0.097656 | 0.048828 | 0 |
---|---|---|---|---|---|---|---|
Replicate 1 | 57809 | 29980 | 15540 | 7861 | 3913 | 1949 | 3 |
Replicate 2 | 58424 | 30336 | 15369 | 7579 | 3948 | 1993 | 3 |
Replicate 3 | 58075 | 29993 | 15535 | 7723 | 3908 | 1999 | 2 |
Replicate 4 | 57953 | 29694 | 15532 | 7798 | 3935 | 2044 | 4 |
Arith. Mean | 58065.25 | 30000.75 | 15494 | 7740.25 | 3926 | 1996.25 | 3 |
S.D. | 262.7171 | 262.665 | 83.39864 | 121.4012 | 18.77942 | 38.86194 | 0.816497 |
3. Cell measurement
0h | 2h | 4h | 6h | |
---|---|---|---|---|
Negative Control (Colony 1) | 0.00475 | 0.0275 | 0.14125 | 0.25875 |
Negative Control (Colony 2) | 0.00975 | 0.0315 | 0.13275 | 0.235 |
Positive Control (Colony 1) | 0.007 | 0.01475 | 0.12125 | 0.25875 |
Positive Control (Colony 2) | 0.00825 | 0.019 | 0.1235 | 0.26925 |
Device 1 (Colony 1) | 0.0045 | 0.00375 | 0.00875 | 0.01 |
Device 1 (Colony 2) | 0.0045 | 0.00325 | 0.009 | 0.007 |
Device 2 (Colony 1) | 0.006 | 0.01825 | 0.12825 | 0.2665 |
Device 2 (Colony 2) | 0.00925 | 0.02325 | 0.141 | 0.26375 |
Device 3 (Colony 1) | 0.0065 | 0.02225 | 0.13425 | 0.2735 |
Device 3 (Colony 2) | 0.0105 | 0.0275 | 0.144 | 0.26325 |
Device 4 (Colony 1) | 0.00375 | 0.00325 | 0.00875 | 0.0445 |
Device 4 (Colony 2) | 0.0065 | 0.004 | 0.0085 | 0.0125 |
Device 5 (Colony 1) | 0.00575 | 0.02375 | 0.14775 | 0.28175 |
Device 5 (Colony 2) | 0.00875 | 0.019 | 0.1275 | 0.27675 |
Device 6 (Colony 1) | 0.006 | 0.02425 | 0.1505 | 0.2855 |
Device 6 (Colony 2) | 0.0085 | 0.02125 | 0.1005 | 0.2625 |
0h | 2h | 4h | 6h | |
---|---|---|---|---|
Negative Control (Colony 1) | 39.5 | 26.25 | 123 | 79 |
Negative Control (Colony 2) | 34.5 | 15 | 120 | 112 |
Positive Control (Colony 1) | 72.5 | 316.25 | 1272.75 | 2534.25 |
Positive Control (Colony 2) | 68.25 | 318.25 | 1269.25 | 2440.25 |
Device 1 (Colony 1) | 557.5 | 563 | 940.5 | 1046.5 |
Device 1 (Colony 2) | 461.75 | 355.25 | 633.75 | 781.25 |
Device 2 (Colony 1) | 83.75 | 374.75 | 2280.25 | 5128.5 |
Device 2 (Colony 2) | 116.25 | 494.25 | 2128.75 | 5528.5 |
Device 3 (Colony 1) | 3 | 29.5 | 154.75 | 232.75 |
Device 3 (Colony 2) | 28.5 | 29.75 | 161 | 258.5 |
Device 4 (Colony 1) | 94 | 172.5 | 327.75 | 769 |
Device 4 (Colony 2) | 71.75 | 165.5 | 286 | 443.5 |
Device 5 (Colony 1) | 7.5 | 196 | 614.5 | 858.25 |
Device 5 (Colony 2) | 14.5 | 189.75 | 796.75 | 987 |
Device 6 (Colony 1) | 60.25 | 29.5 | 129.5 | 164 |
Device 6 (Colony 2) | 33.75 | 26.5 | 97.75 | 165 |
0h | 2h | 4h | 6h | |
---|---|---|---|---|
Negative Control (Colony 1) | 0.045689 | 0.00469 | 0.004282 | 0.001499 |
Negative Control (Colony 2) | 0.018247 | 0.002376 | 0.004481 | 0.002352 |
Positive Control (Colony 1) | 0.055533 | 0.105486 | 0.051542 | 0.048113 |
Positive Control (Colony 2) | 0.04131 | 0.082437 | 0.050482 | 0.044579 |
Device 1 (Colony 1) | 0.607923 | 0.745125 | 0.532061 | 0.520307 |
Device 1 (Colony 2) | 0.522086 | 0.640948 | 0.355477 | 0.553925 |
Device 2 (Colony 1) | 0.070761 | 0.101285 | 0.087415 | 0.094541 |
Device 2 (Colony 2) | 0.063885 | 0.104249 | 0.074675 | 0.103012 |
Device 3 (Colony 1) | 0.001716 | 0.006526 | 0.005688 | 0.004194 |
Device 3 (Colony 2) | 0.01537 | 0.005312 | 0.005489 | 0.004827 |
Device 4 (Colony 1) | 0.118182 | 0.265189 | 0.184017 | 0.084842 |
Device 4 (Colony 2) | 0.054166 | 0.218167 | 0.166671 | 0.177319 |
Device 5 (Colony 1) | 0.00281 | 0.040555 | 0.020409 | 0.014984 |
Device 5 (Colony 2) | 0.007484 | 0.049747 | 0.030945 | 0.017517 |
Device 6 (Colony 1) | 0.050096 | 0.005969 | 0.004225 | 0.002821 |
Device 6 (Colony 2) | 0.024278 | 0.006149 | 0.004835 | 0.003078 |