Introduction
It is significant for synthetic biology to develop a reliable and repeatable measurement, the same to all the other engineering disciplines. We HUST-China have volunteered to test some RBS devices (BCDs) that are intended to make gene expression more precise and reliable by measure the expression level of GFP, in order to help the iGEM community collect data about how reliable will these devices turn out to be in labs around the world.
Interlab Resultn
OD600 Reference Point
Fluorescence standard curve
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.
NOTE: 50uM Sample exceeds the range of measurements
OD600 Reference Point
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.
Final scaling level determined from medium-high points likely to be less impacted by saturation or pipetting error.
If needed, you can shift which points are used, but it is likely better to correct instrument settings and protocol.
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.
Click on the picture and you can see a clearer picture.Then click on the return key of the browser and you can return.