The InterLab Measurement Study is an international effort to increase reproducibility of biomedical research data [1]. Genetically engineered constructs are highly sensitive to environmental factors, and as well as the biological instruments being used. Since minimizing the discrepancies of environmental factors is difficult, the InterLab Study mitigates inconsistencies by focusing on the instruments used. The purpose of this year’s InterLab Study is to address variations in fluorescence measurements, particularly with respect to green fluorescent protein (GFP). | The InterLab Measurement Study is an international effort to increase reproducibility of biomedical research data [1]. Genetically engineered constructs are highly sensitive to environmental factors, and as well as the biological instruments being used. Since minimizing the discrepancies of environmental factors is difficult, the InterLab Study mitigates inconsistencies by focusing on the instruments used. The purpose of this year’s InterLab Study is to address variations in fluorescence measurements, particularly with respect to green fluorescent protein (GFP). |
Revision as of 04:01, 4 October 2017
INTERLAB STUDY
Introduction
Methods
Protocol
Calibrations
Cell Measurements
Detailed Instrument Settings
The fluorescence and absorbance settings are listed below.Flourescence Measurements
GFP IGEM | 4000065 |
---|---|
Top mirror | FITC |
Bottom mirror | N/A |
Exc. filter | FITC 485 |
Using of excitation filter | Top |
2nd exc. filter | N/A |
Using of 2nd excitation filter | Top |
Ems. filter | FITC 535 |
2nd ems. filter | N/A |
Measurement height | 6 mm |
Number of flashes | 10 |
Number of flashes integrated | 1 |
PMT gain | 1 |
Limits of excitation light | 1% |
Range of excitation light | 100% |
Reference AD gain | 2 |
Reference signal | 388381 |
Last edited | 9/21/17 11:34 |
Last edited by | EnVision |
Factory preset | No |
Absorbance 600 settings:
Absorbance @ 600 | 2000005 |
---|---|
Exc. filter | Photometric 600 |
Measurement height | 13 mm |
Number of flashes | 10 |
Number of flashes integrated | 1 |
Limits of excitation light | 100% |
Range of excitation light | 100% |
Reference AD gain | 8 |
Reference signal | 260565 |
Continuous measurement | No |
Last edited | 3/29/16 0:37 |
Last edited by | EnVision |
Factory preset | No |
Filters Used:
FITC 485 | 102 |
---|---|
Filter type | Excitation |
Description | X485 CWL=485nm BW=14nm Tmin=60% |
Used with | Absorbance DELFIA - Time-resolved Fluorescence Fluorescence Intensity Fluorescence Polarization LANCE - Time-resolved Fluorescence |
Last edited | 10/21/10 13:14 |
Last edited by | EnVision |
Factory preset | Yes |
FITC 535 | 206 |
Filter type | Emission |
Description | M535 CWL=535nm BW=25nm Tmin=50% |
Used with | Absorbance DELFIA - Time-resolved Fluorescence Fluorescence Intensity Fluorescence Polarization LANCE - Time-resolved FluorescenceLuminescence DELFIA - Time-resolved Fluorescence Fluorescence Intensity LANCE - Time-resolved Fluorescence |
Last edited | 6/25/07 11:39 |
Last edited by | Installation |
Factory preset | Yes |
Photometric 600 | 319 |
Filter type | Excitation |
Description | P600 CWL=600nm BW=8nm Tmin=40% |
Used with | Absorbance DELFIA - Time-resolved Fluorescence Fluorescence Intensity Fluorescence Polarization LANCE - Time-resolved FluorescenceLuminescence DELFIA - Time-resolved Fluorescence Fluorescence Intensity LANCE - Time-resolved FluorescenceAbsorbance DELFIA - Time-resolved Fluorescence Fluorescence Intensity Fluorescence Polarization LANCE - Time-resolved Fluorescence |
Last edited | 6/25/07 11:39 |
Last edited by | Installation |
Factory preset | Yes |
Results