Cas13 amplification of 'P1-P2-Sp'
Amplification
Kelly
Since it seemed that indeed the right primers were used, it was more likely that #8 from 14/08/2017 was in fact the P1-P2-Spacer insert complete with prefix and suffix. The PCR product at least had the right size. Therefore, this part was amplified in two parallel PCRs.
PCR programme 1: Used primers IG0006 and IG0007. Expected size of product: 4500 bp. Tm used: 63.5°C.
PCR programme 2: Used primers IG0022 and IG0023. Expected size of product: 4350 bp. Tm used: 72°C.
Furthermore, the PCR (Phusion polymerase) protocol was followed.
Results: For both primer sets the sample containing DMSO (lane 3 and 5) gives a more specific result than the sample without DMSO. Primer set IG0006 and IG0007 seems to give a more specific result. Conclusion: sample should be gel extracted.
Gel Isolation
Kelly
The remainder of C45 was loaded on gel (~21uL + 3.5uL loading dye). It was cut from gel and purified. Finally, the concentration was measured on the Nanodrop, yielding a concentration of 4.76 ng/uL. The gel used was an 0.6% agarose gel.
PCR amplificication
Kelly
Both the gel extracted sample and the PCRed samples with and without DMSO and primers IG00006-IG0007 were used as template for another round of PCR. For this PCR, the same programsas before were used (PCR (Phusion polymerase) protocol). We used GC buffer and DMSO for all reactions. The following templates and primers were used:
Sample |
Containing |
Pirmers |
G6 |
Gel extracted "8" (C46) |
IG0006-IG0007 |
C23 |
Gel extracted "8" (C46) |
IG0023-IG0022 |
6+ |
PCRed "8" with DMSO |
IG0006-IG0007 |
6- |
PCRed "8" without DMSO |
IG0006-IG0007 |
The PCR products were loaded on an agarose gel an ran for 45 minutes.
We drew the following conclusions:
Both gel extracted samples barely have a yield and the product is the wrong size.
The samples were very diluted (for "8"), but at least they had a vague of the right size (even though only 0.4 μL was used as template).
All samples were purified (from another gel).
The prodcuts were Nanodropped (DNA concentration measurement (NanoDrop)), which gave th following results:
Sample |
Concentration (ng/μL) |
260/280 |
6- |
141.8 |
1.92 |
6+ |
192.6 |
1.90 |
22- |
237.2 |
1.85 |
22+ |
250.4 |
1.83 |
Sequencing samples
Kelly
The gel extracted "8" and the PCRed "8" +DMSO were sent for sequencing with primer IG0003 (code: 18A1ZAG096) following the sequencing protocol from Macrogen was followed.