Team:NAWI Graz/InterLab

INTERLAB MEASUREMENT STUDY


Reliable and repeatable measurement is key to compare and analyse data from different labs all over the world. To support the effort to create detailed protocols to measure fluorescence and improve the possibility of comparing data, the iGEM Team NAWI_Graz 2017 decided to participate in the Fourth International InterLaboratory Measurement Study in synthetic biology. The goal is to establish a GFP measurement protocol based on engineering principles.

Our tasks were to follow exactly the iGEM Plate Reader Protocol , fill in the prepared excel sheets and send the results to the iGEM headquarter.

Results

For this study we needed to transform 8 devices into the presupposed bacteria strain E.coli DH5&alpha.
This included a positive control BBa_I20270
the constitutive tetR repressible promoter as negativ control BBa_R0040
Test Device 1 BBa_J364000
Test Device 2 BBa_J364001
Test Device 3 BBa_J364002
Test Device 4 BBa_J364003
Test Device 5 BBa_J364004
and Test Device 6 BBa_J364005
They are all stored in pSB1C3 and were cultured on LB-agar plates with a chloramphenicol concentration of 100 µg/ml. As we had grown colonies the day after the transformations, we could start with the cell measurement protocols.

Calibration Protocols

1. OD600 Reference point
Results of OD600 measurement of 1 ml LUDOX and 1 ml H2O with dH2O as blank. LUDOX Table
Fig. 1: OD600 measurement
2. Protocol fluorescein fluorescence standard curve:
fluorescein Table
Fig. 2: Fluorescein fluorescence plate reader measurement with gain 48 & 70 in a 96 well plate. The concentration cuts in half from pure fluorescein in line 1 to 9.766*10-4 µl fluorescein diluted in PBS in line 11 and PBS only in line 12. All volumes are 100 µl.

The task was to find the optimal plate reader settings for the cell measurement on Day 3 through a fluorescein dilution in PBS and measuring its flourescence (excitation 485 nm, emission 531 nm). We came up with the problem, that the same gain settings could not be used for both, fluorescein and cell measurement as in each case one of the measurements only displayed 50% usable data. Additionally, only a gain of 48 or less displayed a complete set of fluorescein values. But the producing company of our plate reader only recommends a gain range for 50-150. So not even gain 48 displays “usable" values. We found the possibility for wider detection ranges in the product help documents, but our plate reader model Synergy MX from BioTek did not offer this option. Problems with the dilution by pipetting can be excluded.

[Fluorescein standard curve]
[Standard curve (log)]
Fig. 3 & 4: This diagrams display the amount of fluorescence according to the fluorescein µM concentration diluted in PBS at a total volume of 100 µl per well.

As values bigger than 100 000 are labeled as OVERFLOW and do not represent a usable number, the excel algorithm could not display them in usable curves.


Cell Measurement Protocol

Day 1 OD600 Measurement

Day 2 Transformation

The standard dilution sheet only calculates volumes for a target of 10 ml. The official InterLab protocol wants a target volume of 12 ml but also the use of the dilution calculation sheet, so we made a decision and sticked clearly to the protocol and did not decide to change the sheet algorithm.

Day 3 Cell growth, sampling and assay. Preparing and taking probes at T0h, T2h, T4h, T6h accordingly to the dilution sheet and plate reader protocol.
Flourescence table
Fig. 5: Raw readings of fluorescence measurement of the day 2 ONC at 0 hours, 2 hours, 4 hours and 6 hours.
As you see Device 1 had the strongest fluorescence during the experiment. Even though we tried several different gain settings. The actual gain we should use, according to our results following the protocol, would be a gain of 48, but most of the devices did not fluorescent enough to be detected with this low gain. Even though we needed so much time to get the best cell measurement settings that T0 grew to an OD600 0.05 by accident, the actual problem stayed the same. Our cells did not grow fast enough to get usable values of the fluorescence measurement at the same setting as the fluorescein measurement produced usable data. The trouble shooting did not bring up good possibilities because the growth settings were not comparable with our project experiments. The fluorescein measurement was done three times with different, but similar buffers. All produced nearly the same data. Maybe you need other plate reader models get better results.
Flourescence table
OD600 measurement
Fig. 6 & 7: This diagrams display the average of the device's fluorescence and OD600 according to the time the probes were taken.

Even though device 2 shows the strongest fluorescence, device 4 grew better resulting in a higher OD600. Even though, nearly all devices and both controls grew in equal speed and reached the same cell density.


Measurement Procedure Details

96 well plate, black, flat, transparent bottom
Read: Fluorescence Endpoint
Excitation: 485 Emission: 531
Band width: 20.0
Optics: Top, Gain: 70 Read speed: Normal, Delay: 100 ms
Measurements/Data Point: 10 Read Height: 8 mm
Temperature: 28,8 °C (no preset)