Team:CCU Taiwan/Results

No Sidebar - Helios by HTML5 UP

Lactate detect

Biobrick Assembly


Figure 1 1%Agarose, 100V. Check the size of insert.
Length: pUCIDT(2705 bp)- J23118-LldR(1006bp)
Length: pUCIDT(2705 bp)- O1-J23118-O2-GFP(1235bp)
Sample 1: pUCIDT-J23118-LldR without digestion
Sample 2: pUCIDT-J23118-LldR digested with EcoRI and PstI
Sample 3: pUCIDT-O1-J23118-O2-GFP without digestion
Sample 4: pUCIDT-O1-J23118-O2-GFP digested with EcoRI and PstI
After received our IDT plasmid DNA, the most important thing was to check the insert size was correct or not. We digested (EcoRI-PstI) it to check the size of plasmid. From Agarose gel result, we knew the length of inserts are correct.


Figure 2 1%Agarose, 100V. Check the size of pSB1A2.
Length: pSB1A2(2079 bp)-E0032(762 bp) digested with EcoRI and PstI
Sample 1: pSB1A2-E0032 without digestion
Sample 2: pSB1A2-E0032 digested with EcoRI and PstI
From these agarose gel we known the size of backbone that we wanted was right. After knowing that, it is the time to assamble them together !!


Figure 3 1% Agarose, 100V. Check the size of assemble [pSB1A2 (2079 bp)]-[O1-J23118-O2-GFP-J23118-LldR (2241 bp)].
Sample 1: [pSB1A2 (2079 bp)]-[O1-J23118-O2-GFP-J23118-LldR (2241 bp)] digested with EcoRI and PstI
Sample 2: [pSB1A2 (2079 bp)]-[O1-J23118-O2-GFP-J23118-LldR (2241 bp)] digested with EcoRI and PstI
Sample 3: [pSB1A2 (2079 bp)]-[O1-J23118-O2-GFP-J23118-LldR (2241 bp)] digested with EcoRI and PstI

Lactate detect function test

After finished assambling our parts, we did some function test, to find the relationship between GFP expression and lactate level.


Figure 4 Fluorescence of each lactate concentration depends on time.
We found that even though fluorescence in each concentration of lactate are floating, the spacing of each data are stable expect data 10 mM. It seems to be saturated when lactate concentration reaches 8 mM.

  1. (a)The fluorescence of each lactate concentration depends on time.

Figure 5 Fluorescence of different time depends on lactate level.
The interval 120 minute ≤ time ≤210 minute, the fluorescence of each concentration of lactate is similarity. Positive correlation exists between fluorescence and concentration of lactate on an interval 120 minute ≤ time ≤270 minute.

  1. (b)Fluorescence intensity of different time period depends on lactate level.

Figure 6 Combine Figure 4 and Figure 5 together.
We found that fluorescence level was high at the beginning. The general trend of fluorescence became pronounced and stable as time goes by.

  1. (c)It’s quite interesting that the fluorescence value always be very high at the beginning but fell as time passed by, even in negative control. The tendency of whole line is similar. And if we just see a single time point (b), a highly positive correlation between fluorescence and lactate level can be observed.

From figure 4~6, we found that there is a positive correlation between lactate concentration and fluorescence value. Below we took several points from 210 min and 270 min to create figures.


Figure 7 Fluorescence vs. Concentration of lactate at time=210 min